Complications Leading to Death in Patients Supported by the Impella 5.5: Analysis From the FDA MAUDE Database
database[Title] 2025-05-14
Summary:
CONCLUSIONS: Cardiac perforation emerged as the most frequent complication associated with Impella 5.5-related deaths, followed by device malfunction. These findings highlight the need for improved operational techniques, better device reliability, and refined safety protocols for Impella 5.5 usage.